Differential Pressure Sensor Transmitter ±100pa ±500pa ±1Kpa 4-20mA Hart NPT1/4 Monocrystalline Silicon Pressure Transmitter 1Pcs(0-50Kpa)

Imported from USA | Ships in 10 working days

5% instant discount on non COD

1

₹ 225300/-

The price at cart2india.com is inclusive of custom duties, GST, freight forwarding and other charges.

Free Shipping across India. Powered by BlueDart

Product Information


Specification
Brand : FRPKPNOX
BulletPoint : Pressure Transmitters
Color : Output 4-20mA Hart
ExternallyAssignedProductIdentifier : 7341569239948
ItemName : Differential Pressure Sensor Transmitter ±100pa ±500pa ±1Kpa 4-20mA Hart NPT1/4 Monocrystalline Silicon Pressure Transmitter 1Pcs(0-50Kpa)
ItemPackageDimensions_Height : 1 centimeters
ItemPackageDimensions_Length : 3 centimeters
ItemPackageDimensions_Width : 2 centimeters
ItemPackageQuantity : 1
ItemTypeKeyword : pressure-transmitters
Manufacturer : FRPKPNOX
MeasurementAccuracy : 1
ModelNumber : FRPKPNOX
NumberOfBoxes : 1
NumberOfItems : 1
PartNumber : FRPKPNOX
ProductDescription : Pressure Transmitters
Product Details:
1. Pressure Range: -1 ~ 400Bar . Customzied
2. Output Signal : 4-20mA +Hart Default. Supply Rs485 . Contact me for check.
3. Medium Temperature : -20~60
4. Material: 304SS Thread. Low copper aluminum alloy shell. Epoxy coating layer.
5. Connected Thread: M20*1.5 Default. Contact me if you need others.
Key description and use:
Z (zero) S (Full) M (function key)
1. Press the button to unlock
Press Z and S keys at the same time for more than 5 seconds to unlock (LCD screen display: OPEN)
2, key zero
Apply ZERO pressure to the transmitter, press the S key for about 5 seconds, the transmitter output 4.000MA current, complete the zero operation (LCD display: ZERO)
3, button full
Apply full point pressure to the transmitter, press the S key for about 5 seconds, the transmitter output 20.000MA current, complete the full adjustment operation (LCD display: SPAN)
4. Clear PV value to zero
Put the transmitter directly at atmospheric pressure, press the button to unlock, and then press the Z and S keys at the same time for about 5 seconds, you can set the current PV value to 0, (LCD PV=0)
Note: If the current PV value deviates from the 0 value by more than 50%FS, the PV value is invalid
5, screen display setting unit
Kpa, atm, MPa, bar, mbar, pa, psi, mmHO, mmHG, gcm, kgcm
Model Number : QD-3051-VD
Max Measuring Range : 400 Bar & Above
Origin : Mainland China
Name : 3051 Monocrystalline Silicon Pressure Sensor Transmitter
Range : -1~400bar
Output Signal : 4-20mA + Hart
Power Supply : DC24V
Connected Thread : M20*1.5
Accuracy : 0.075%